Please use this identifier to cite or link to this item: https://hdl.handle.net/10216/154780
Author(s): Moller, P
Azqueta, A
Rodriguez-Garraus, A
Bakuradze, T
Richling, E
Bankoglu, EE
Stopper, H
Bastos, VC
Langie, SAS
Jensen, A
Ristori, S
Scavone, F
Giovannelli, L
Wojewodzka, M
Kruszewski, M
Valdiglesias, V
Laffon, B
Costa, C
Costa, S
Teixeira, JP
Marino, M
Del Bo', C
Riso, P
Zheng, CY
Shaposhnikov, S
Collins, A
Title: Long-term cryopreservation of potassium bromate positive assay controls for measurement of oxidatively damaged DNA by the Fpg-modified comet assay: results from the hCOMET ring trial
Publisher: Oxford University Press
Issue Date: 2023
Abstract: The formamidopyrimidine DNA glycosylase (Fpg)-modified comet assay is widely used for the measurement of oxidatively generated damage to DNA. However, there has not been a recommended long-term positive control for this version of the comet assay. We have investigated potassium bromate as a positive control for the Fpg-modified comet assay because it generates many Fpg-sensitive sites with a little concurrent generation of DNA strand breaks. Eight laboratories used the same procedure for the treatment of monocytic THP-1 cells with potassium bromate (0, 0.5, 1.5, and 4.5 mM) and subsequent cryopreservation in a freezing medium consisting of 50% foetal bovine serum, 40% RPMI-1640 medium, and 10% dimethyl sulphoxide. The samples were analysed by the Fpg-modified comet assay three times over a 3-year period. All laboratories obtained a positive concentration-response relationship in cryopreserved samples (linear regression coefficients ranging from 0.79 to 0.99). However, there was a wide difference in the levels of Fpg-sensitive sites between the laboratory with the lowest (4.2% Tail DNA) and highest (74% Tail DNA) values in THP-1 cells after exposure to 4.5 mM KBrO3. In an attempt to assess sources of inter-laboratory variation in Fpg-sensitive sites, comet images from one experiment in each laboratory were forwarded to a central laboratory for visual scoring. There was high consistency between measurements of %Tail DNA values in each laboratory and the visual score of the same comets done in the central laboratory (r = 0.98, P < 0.001, linear regression). In conclusion, the results show that potassium bromate is a suitable positive comet assay control.
DOI: 10.1093/mutage/gead020
URI: https://hdl.handle.net/10216/154780
Source: Mutagenesis. 2023 Oct 14;38(5):264-272. doi: 10.1093/mutage/gead020.
Document Type: Artigo em Revista Científica Internacional
Rights: restrictedAccess
Appears in Collections:ISPUP - Artigo em Revista Científica Internacional

Files in This Item:
File Description SizeFormat 
mollerazqueta2023-arci-r.pdf
  Restricted Access
677.22 kBAdobe PDFView/Open
mollerazqueta2023-arci-r-com.pdf
  Restricted Access
645.69 kBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.