Please use this identifier to cite or link to this item:
https://hdl.handle.net/10216/70131
Author(s): | Miguel V. Correia S. Cunha D. Viegas C. Cunha P. Diniz E. Mendes |
Title: | Acquisition system for data analysis of gait with lower limb prostheses - ANDAR |
Issue Date: | 2005 |
Abstract: | This paper describes the implementation of a gait analysis device used to measure the use of prosthesis by lower limb amputated subjects. Based on dead-reckoning sensors only (two accelerometers, one gyroscope and pressure sensors, all solid state), it is used to compute the travelled distance, number of steps, motion speed, leg aperture and related parameters from which assessment of comfort and patient adaptation to the prosthesis can be derived. The core of the real-time data processing set of algorithms is a Kalman filter, as here described, together with some results. |
Subject: | Engenharia electrotécnica, electrónica e informática Electrical engineering, Electronic engineering, Information engineering |
Scientific areas: | Ciências da engenharia e tecnologias::Engenharia electrotécnica, electrónica e informática Engineering and technology::Electrical engineering, Electronic engineering, Information engineering |
URI: | https://repositorio-aberto.up.pt/handle/10216/70131 |
Source: | Proceedings of IADAT International Conference on Automation, Control and Instrumentation |
Document Type: | Artigo em Livro de Atas de Conferência Internacional |
Rights: | openAccess |
License: | https://creativecommons.org/licenses/by-nc/4.0/ |
Appears in Collections: | FEUP - Artigo em Livro de Atas de Conferência Internacional |
This item is licensed under a Creative Commons License